Pannoramic 250 FLASH II 1.18.1 User's Guide
Manual Exposure
By selecting the Use manual exposure times, you can determine the exposure times. The
values are typically set to Auto, meaning that the exposure times are determined
automatically. After selecting the corresponding filter and activating the live image, you can
preset the Exposure value.
It is recommended to go to various positions in order to determine the intensity distribution
throughout the specimen and define a representative value. You can accept the value for
presetting by clicking Set Value. Repeat the procedure for all other filters.

4.5.5 Service panel

This panel contains control and maintenance functions.
Emergency Manual Slide Removal Position
The specimen holder will be moved to a position where the slide can be removed manually
after opening the cover (see section 5.3.1 Opening the protective cover).
Slide Removed Manually
Click this button to notify the software that you have removed the slide manually.
Show images during scanning
If you select this option, you will be able to see the currently scanned image field during
scanning in the live image display field. This slows down digitization process and should be
activated only with purpose.
New Compensation Image for Preview Camera
By clicking this button, you can create compensation image for the preview camera if it fails
to display a satisfactory preview image.
May 29, 2014 – Rev. 1
Figure 67 – Service panel