Pannoramic 250 FLASH II 1.18.1 User's Guide
Brightfield Compensation
The Use saved compensation image function enables to use previously recorded reference
images for all slides. Use this function when the slide is completely covered by the specimen
(for example, blood smear) thereby making compensation image capture impossible. For
such slides, use an empty slide with coverslip to grab a compensation image (for more
information, see section 4.2.2. Image Compensation / Brightfield).
If you select this option, the software will not create new compensation images for each
slide. Compensation images are reference images used for white balancing and shading.
The Scanned slide destination field determines where to digitize the virtual slides.
Select the Scan To CaseCenter function to scan directly to a dedicated CaseCenter server.
For information on how to use the ScanToServer application, see section 4.7 Scan To Server.
If you enable the Directory function, the virtual slides will be digitized in the specified
folder. You can browse for the folder with the Browse button. If you specify a folder that
does not exist, the software will create it.
All parameter inputs are saved for the next software session.