Pannoramic MIDI 1.18 User's Guide
With the Z-stack method separate layers are created and stored, and can be opened with
Pannoramic Viewer. The Z-stack slide contains the extended focus layer. Both for Extended
focus and Z-Stack scanning methods, you can determine the desired number of focus levels
with the Focus levels selection box. The current focus value is the starting point. If you set 5
for example, one current, two focus levels above and below FOVs are captured. Therefore, it
is advised to choose an odd number for the focus levels.
Focus step size determines the spacing between focus levels. One step is around 0.2 µm.
Default settings are: Number of focus levels: 7; Focus step size: 5.

4.3.6 Service panel

This panel contains control and maintenance functions.
Emergency Manual Slide Removal Position
The specimen holder will be moved to a position where the slide can be removed manually
after opening the cover (see section
Slide Removed Manually
Click this button to notify the software that you have removed the slide manually.
April 3, 2014 - Rev 1
Figure 39 – Focus level and step size
Figure 40 – Service panel
5.3.1 Opening the protective
3DHISTECH Ltd.
4 Software Description
covers).