Pannoramic 250 FLASH II 1.18.1 User's Guide
When in Automatic Mode, the form contains 25 rows for all the slots per magazine
(Figure 80). To additional magazines click Add+ panel next to the Magazine 01 panel.
May 29, 2014 – Rev. 1
Figure 80 – Scan To Server window in Automatic Mode
To scan a series of slides to the server (within a range), give the number of the
starting (First slide) and the final slide (Last slide) in the appropriate field. For
example, if you want to scan slides to server from position 3 to 14 of the magazine,
type: First slide: 3; Last slide: 14.
There are three options at the bottom of the window.
1. Server side barcode parsing option can be selected if CaseCenter is connected
with the supported HIS/LIS. This way the barcode read by the software is being
identified, and the predefined location of that specific case in CaseCenter is
automatically selected as scanning destination based on the CaseID (containing
slide and patient data stored on HIS/LIS). Upon activating this option, Slide
name, Folder, Case, and Block field data cannot be modified, only a profile can
be selected.
2. The Generate slide name if no name was defined option can be activated when
you do not want to type in a unique name for each slide.
3. Upon activating the Read barcode if no name is defined on the list option, the
slide barcode will be read and set in the Barcode field.
Use masks for automatic fills. Activate the Use mask option and the mask fields
become fillable. To save masks, click the Save Mask button. Click Set Mask to
modify masks. For example, by creating a string for the slide name fields based on
the characters of the barcode, a unique name will be generated.