Pannoramic 250 FLASH II 1.18.1 User's Guide
May 29, 2014 – Rev. 1
Filter name
The name of the filter is displayed here.
Exposure Settings
Exposure time – You can set the fix exposition time here. By selecting the
Auto button, the exposition time is set to
calculated by the software). If the fix position is set, the Exposure time
weight value is not valid.
You can set the Exposure time weight (%) here (displayed under TW in the
Filter Selection frame).
In full automatic mode, the system automatically calculates the exposition
time to the given channel. You can increase / decrease this amount here in a
range of 50%-200%.
Multilayer Settings
Enable Multilayer function and the concerning settings can be activated in
this field, accordingly, you can set the Extended focus and Z-stack values for
each channel separately.
Digital Gain
Note:
This function is camera-dependent, works only with
AxioCam MRm Rev 3 camera (Index value ranges from 0
to 4) and pco.edge 4.2 (Index values: 0-6).
You can enter the Index value by entering it in the box or by selecting the
value with the arrow buttons. If you raise the Index value, the exposition
time is halved by every step, and the image becomes noisier.
You can move between the filters by clicking Previous Filter or Next Filter.
3DHISTECH Ltd.
4 Software Description
(the exposition time is
<AUTO>