Replacing The Hxp 120 Lamp Module (For Fluorescent Option) - 3D Histech Pannoramic 250 FLASH II User Manual

Table of Contents

Advertisement

Pannoramic 250 FLASH II 1.18.1 User's Guide

5.3.3 Replacing the HXP 120 Lamp Module (for Fluorescent Option)

This section recites some extracts from the HXP 120 operating manual. For detailed information,
especially instructions regarding safety, consult that manual. Also, no alignment is required for the
lamp module. Accordingly, the module is fully operational after lamp replacement.
Hot surface! Wait at least twenty minutes to allow the lamp surface to
cool down sufficiently.
1. Disconnect power plug (separate all poles from the mains).
2. Unfix screws of lateral cover plate and remove it from the unit housing (see Figure 89).
3. Press plug securing latches inward (1) then disconnect plug of the lamp module (2) (see
Figure 90).
May 29, 2014 – Rev. 1
Warning!
Figure 89 – Removing the lateral cover of the HXP 120
Figure 90 – Disconnect plug