Pannoramic 250 FLASH II 1.18.1 User's Guide
With the Z-stack method separate layers are created and stored, and can be opened with
Pannoramic Viewer. The Z-stack slide contains the extended focus layer.
Both for Extended focus and Z-Stack scanning methods, you can determine the desired
number of focus levels with the Focus levels selection box. The current focus value is the
starting point. If you set 5 for example, one current, two focus levels above and below FOVs
are captured. Therefore, it is advised to choose an odd number for the focus levels.
Focus step size determines the spacing between focus levels. One step is around 0.2 µm.
The default settings are: Number of focus levels: 7; Focus step size: 5.

4.3.6 Service panel

This panel contains control and maintenance functions.
Emergency Manual Slide Removal Position
The specimen holder will be moved to a position where the slide can be removed manually
after opening the cover (see section 5.3.1 Opening the protective cover).
Slide Removed Manually
Click this button to notify the software that you have removed the slide manually.
Show images during scanning
If you select this option, you will be able to see the currently scanned image field during
scanning in the live image display field. This slows down digitization process and should be
activated only with purpose.
New Compensation Image for Preview Camera
By clicking this button, you can create compensation image for the preview camera if it fails
to display a satisfactory preview image.
May 29, 2014 – Rev. 1
Figure 49 – Service panel