Pannoramic 250 FLASH II 1.18.1 User's Guide
General threshold options
Remove coverslip – If selected, the edges of the coverslip are not detected as specimen to
•
be scanned. Depending on the threshold level, it is possible that the edges of the coverslip
are detected as areas to be digitized.
It requires lots of time (since it cannot find focus neither on the
edges, nor on specks) to digitize coverslip edges. Select Remove
coverslip function to eliminate those areas.
Do not use Remove coverslip function on TMA slides! The layout of
TMA slides may mislead the detecting algorithm.
Fill holes in scan area – If selected, the holes inside the specimen will also be scanned.
•
Often these holes contain small sample objects that otherwise would not be scanned.
Remove specks – If selected, the scanner does not digitize objects smaller than the
•
predetermined size. This helps filtering specks and other uninteresting objects out. Set the
maximum size of specks not to be digitized by adjusting the Speck size slider.
Scan area dilation – Adjust the slider To enlarge or decrease the scanned area. The edge is
•
extended discretely by 100 µm each time. The recommended value for this setting is 1000
μm in Flash scanning mode.
May 29, 2014 – Rev. 1
Figure 37 – Preview image with and without coverslip edge
Caution!
Warning!
3DHISTECH Ltd.
4 Software Description
59(137)