Pannoramic 250 FLASH II 1.18.1.1 User's Guide
4. Move the lever (1) to the open position, which is to the right when facing the front of the unit
(see Figure 9).
5. Attach the Lumencor unit so the contact surfaces at the mounting point are tightly
compressed (see Figure 10 – Step 1).
6. Secure the locking lever by moving it to the left (see Figure 10 – Step 2). If the lever stucks at
around the center position, apply a bit more pressure to the Lumencor unit.
7. Close the protective cover of the Pannoramic 250 FLASH II.
8. Connect power supply cables both of the scanner and the illumination unit.
9. Connect TTL trigger cable between pco.edge 4.2 camera (out 4) and the Lumencor unit (TTL),
and USB-to-RS232 communication cable between the Lumencor unit and the computer.
May 29, 2014 – Rev. 1
Figure 9 – Positioning locking lever
Figure 10 – Mounting the Lumencor unit
3DHISTECH Ltd.
2 Installation
27(137)