Solvent Information

Observe the following recommendations on the use of solvents.
• Brown glass ware can avoid growth of algae.
• Avoid the use of the following steel-corrosive solvents:
• Avoid the use of dimethyl formamide (DMF). Polyvinylidene fluoride
320
solutions of alkali halides and their respective acids (for example,
lithium iodide, potassium chloride, and so on),
high concentrations of inorganic acids like sulfuric acid and nitric acid,
especially at higher temperatures (if your chromatography method
allows, replace by phosphoric acid or phosphate buffer which are less
corrosive against stainless steel),
halogenated solvents or mixtures which form radicals and/or acids, for
example:
→ 2COCl
2CHCl
+ O
3
2
This reaction, in which stainless steel probably acts as a catalyst, occurs
quickly with dried chloroform if the drying process removes the
stabilizing alcohol,
chromatographic grade ethers, which can contain peroxides (for
example, THF, dioxane, diisopropyl ether) should be filtered through dry
aluminium oxide which adsorbs the peroxides,
solvents containing strong complexing agents (e.g. EDTA),
mixtures of carbon tetrachloride with 2-propanol or THF.
(PVDF), which is used in leak sensors, is not resistant to DMF.