Troubleshooting and Diagnostics
Troubleshooting Guide for the Sample Transport Assembly

Poor alignment

With Error messages
• motor overtemp (0 or 2 or 3)
• movement failed (0 or 2 or 3)
Personal damage, damage to the module
WA R N I N G
When a motor over temperature message has occurred, the sampler must be turned OFF
N O T E
for about 10 min to allow the motor to cool down.
1 Start automatic referencing in Lab Advisor. For this procedure it is
2 Exchange the gripper arm assembly (Gripper assembly (G1313-60010)), see
3 Exchange Transport assembly ( Sample Transport Assembly
4 Exchange the Autosampler Main Board ( Autosampler Specific Board
5 For all three steps (step 2, 3, and 4) you have to do an automatic referencing
160
Some of these procedures require a trained service engineer. Persons who are not
qualified MUST NOT perform these procedures.
mandatory to use the calibration tool.
On completion of the procedure, the corrected values are stored in the
instrument firmware.
"Exchange the Gripper Arm"
(G7129-60600)).
(G7129-65800)) or the fusion board ( Fusion Board (G4200-65880)).
again to store the correct values in the instrument firmware again.
on page 216.