Valve switch not found.

Error ID: 34319
The injection valve has failed to switch into the position. The valve position
sensor monitors the position of the valve. If the valve fails to switch into the
position (mainpass (main path) or bypass), or if the sensor fails to recognize
the valve position, the error message is generated.
Probable cause
1
2

Metering Initialization Failed

Error ID: 34322
The metereing device has failed to do the initialization. If the piston fails to
move to the home position, or if the metering device is not recognized by the
instrument.
Probable cause
1
2
3
4
5
Agilent InfinityLab LC Series Vialsamplers User Manual
Defective rotor seal/stator head.
Defective injection valve.
Metering device is not recognized.
Loose belt.
Rear wheel in the hydraulic unit is loose.
Defective metering-drive motor.
Defective metering device
Error Information
Sampler Error Messages
Suggested actions
Exchange the rotor seal/stator head.
Please contact your Agilent service
representative.
Suggested actions
Configure the metering device in LabAdvisor
Please contact your Agilent service
representative.
Please contact your Agilent service
representative.
Please contact your Agilent service
representative.
Please contact your Agilent service
representative.
8
179