Maintenance and Repair

Exchange the Wash Port Assembly

Exchange the Wash Port Assembly

When
Defective flapper
Parts required
p/n
G7129-60033
Preparations
Finish any pending acquisition job and in order to avoid leaks, stop the pump running and remove
the tubings from the solvent bottles. If available close the shutoff valves.
1
In the Local Controller start the maintenance mode and
select Change needle/seat function.
OR
In the Agilent Lab Advisor software select Service &
Diagnostics in the system screen (Tools) Maintenance
Positions > Change Needle/Loop, click Start and wait
until the needle assembly is in maintenance position.
2
Open the front door.
222
Description
Saftey wash port (needle station)
3
Check if the needle is lifted up.
Agilent InfinityLab LC Series Vialsamplers User Manual