8
Error Information
Integrated Column Compartment (ICC) Heater Error Messages
ICC Heater Temp Sensor Failure
Error ID: 30821, 30823,30825
The Integrated Column Compartment (ICC) Heater has detected a reading
issue with the heater sensors in the heating element.
Probable cause
1
2
ICC Heater Over Temp Failure
Error ID: 30824
The Integrated Column Compartment (ICC) Heater has detected a
temperature reading above 100 °C. This exceeds the limit of the heating
element.
Probable cause
1
2
184
Defective cable or connector
Defective ICC heater sensor
Defective cable or connector
Defective ICC heater
Suggested actions
Please contact your Agilent service
representative.
Replace the ICC heater.
Suggested actions
Please contact your Agilent service
representative.
Replace the ICC heater.