Icc Heater Temp Sensor Failure; Icc Heater Over Temp Failure - Agilent Technologies InfinityLab LC Series User Manual

Vialsamplers
Hide thumbs Also See for InfinityLab LC Series:
Table of Contents

Advertisement

8
Error Information
Integrated Column Compartment (ICC) Heater Error Messages

ICC Heater Temp Sensor Failure

Error ID: 30821, 30823,30825
The Integrated Column Compartment (ICC) Heater has detected a reading
issue with the heater sensors in the heating element.
Probable cause
1
2

ICC Heater Over Temp Failure

Error ID: 30824
The Integrated Column Compartment (ICC) Heater has detected a
temperature reading above 100 °C. This exceeds the limit of the heating
element.
Probable cause
1
2
184
Defective cable or connector
Defective ICC heater sensor
Defective cable or connector
Defective ICC heater
Suggested actions
Please contact your Agilent service
representative.
Replace the ICC heater.
Suggested actions
Please contact your Agilent service
representative.
Replace the ICC heater.