Installation of the Optional Sample Cooler or Thermostat
Using the Sample Cooler or Thermostat in an Infinity II Sampler
Control
The Thermostat section of the Control dialog box is available when you have a
sample cooler or sample thermostat installed. It allows you to switch the
cooler or thermostat on and off manually.
Select On to switch on the cooler/thermostat. Specify the required
temperature in the adjacent field. Note that the specified temperature must be
at least 5 °C below ambient for proper temperature control.
Select Off to switch off the cooler/thermostat.