Sample Cooler Error Messages

Sample Cooler Error Messages

Cooler Voltage Failure

Error ID: 30722
The sample cooler board has detected a voltage failure.
Probable cause
1
2

Cooler PCB Error State

Error ID: 30724, 30275
The sample cooler PCB board seems to be defective.
Probable cause
1
2
180
Blown fuses, defective cable
Voltage issue in the sample cooler
Blown fuses, defective cable
Defective PCB boards
Suggested actions
• Check the power cable from the ALS to the
sample cooler.
• Please contact your Agilent service
representative.
Please contact your Agilent service
representative.
Suggested actions
• Check the power cable from the ALS to the
sample cooler.
• Please contact your Agilent service
representative.
Replace the sample cooler.