Exchange the Sample Loop Assembly

Exchange the Sample Loop Assembly

When
When sample loop is visibly damaged, blocked or leaks.
Tools required
p/n
8710-1924
Parts required
p/n
G7129-60500
G7129-60400
G7129-60300
Preparations
Finish any pending acquisition job and in order to avoid leaks, stop the pump running and remove
the tubings from the solvent bottles. If available close the shutoff valves.
Risk of injury by uncovered needle
WA R N I N G
An uncovered needle is a risk of harm to the operator.
When the instrument setup has changed, configure the new setup in the online view of
N O T E
the Chromatographic Data System. See
202
Description
Open-end wrench 14 mm
Description
Sample Loop 100 µL
Sample Loop 40 µL
Sample Loop 20 µL
Do not open the safety cover of the needle station during normal operation.
Wear safety goggles and safety gloves when removing the needle assembly.
Table 12
on page 120.
Agilent InfinityLab LC Series Vialsamplers User Manual