Sample Cooler Error Messages

Cooler Condensate Error

Error ID: 30702, 30709, 30710
The sample cooler has detected condensate inside of the leak pane.
Probable cause
1
2
3
182
Blocked drainage
Defective condensate sensor
Defective leak sensor
Suggested actions
Switch off the instrument and remove the
condensate tubings. Place a beaker underneath
the waste outlet. Check the waste container.
There should be no kinked and sealed tubing.
Switch the sample cooler off/on and check if
after 8 min the error shows up again.
Replace the sample cooler.