Install the Column Shelf

Install the Column Shelf

Tools required
p/n
8710-0510
Preparations
• Switch off the pump
• Switch off the sampler
• Open the doors of the sampler
• Open the ICC
Fire or explosion due to missing leak drainage
WA R N I N G
Solvents may run into the module if Column Shelf or Integrated Column
Compartment are not installed.
1
Remove the capillary connection from the integrated
column compartment to the module.
66
Description
Open-end wrench 1/4 — 5/16 inch
Install either Column Shelf or Integrated Column Compartment before starting the
module.
2
Remove the column.