Install the Sample Cooler or Thermostat

Parts required
#
1
1
OR
1
1
1
1
Preparations
Sampler is installed in the stack.
Flammable refrigerant
WA R N I N G
Formation of flammable gas-air mixtures inside the sample thermostat and
laboratory.
Flammable refrigerant used
WA R N I N G
Agilent InfinityLab LC Series Vialsamplers User Manual
p/n
G7167-60005
G7167-60101
5067-6208
G7167-90171
Keep open fire or sources of ignition away from the device.
Ensure a room size of 4 m
sample thermostat).
Ensure adequate ventilation: typical air exchange of 25 m
floor area.
Keep all ventilation openings in the enclosure clear of obstructions. Do not block
the openings on the circumference of the sample thermostat.
When handling, installing and operating the Sample Thermostat, care should be
taken to avoid damage to the refrigerant tubing or any part of the Sample
Thermostat.
Installation of the Optional Sample Cooler or Thermostat

Install the Sample Cooler or Thermostat

Description
Sampler
Sample cooler
InfinityLab Sample Thermostat
Power cord
Condensate Drainage Kit
Installation of the Infinity II Cooler/Thermostat Condensate
Drainage Tubing Kit
3
3
(1 m
for every 8 g of R600a refrigerant inside of the
3
2
/h per m
of laboratory
4
91