Remove a Column (ICC)

Remove a Column (ICC)

Tools required
p/n
8710-0510
Parts required
p/n
5067-5957
5067-5966
Preparations
• Switch off the pump
• Switch off the sampler
• Open the doors of the sampler
1
Push the latch.
3
Remove the column.
76
Description
Open-end wrench 1/4 — 5/16 inch
Description
Column (up to 30 cm length)
InfinityLab Quick Connect Assy ST 0.12 mm x 105 mm
InfinityLab Quick Turn fitting
2
Open the flap 90 °.
4
Unscrew the right fitting.
Agilent InfinityLab LC Series Vialsamplers User Manual