Tests
ICC Heater Test
The Oven Test is used to evaluate the heating performance of the heating
element. The heating rate over a 10 K interval from the start temperature is
determined. The start temperature must be between 30 °C and 50 °C, and is
determined as follows:
• If the current oven temperature is below 30 °C, the oven tries to attain a
• If the current oven temperature is above 30 °C but below 50 °C, the current
• If the current oven temperature is above 50 °C, an error message is
Sample Cooler Function Test
The Sample Cooler Function Test can be used as a simple verification that the
sample cooler is functioning.
After the test has been started, it begins to acquire data from the cooler's
PT1000 temperature sensor. As soon as the temperature has equilibrated (that
is, the temperature does not change by more than 0.5 °Cover a 10 s period) the
cooler is turned on and measurement is started.
For the test to succeed, three temperature checkpoints must be reached within
a specified time.
Agilent InfinityLab LC Series Vialsamplers User Manual
temperature of 30 °C. 30 °C is used as the start temperature.
oven temperature is used as the start temperature.
displayed. The oven must then be allowed to cool to below 50 °C before the
test can be run.
Troubleshooting and Diagnostics
7
Tests
155