Preparing the Module

For best performance of the module
• When using the module in a system with a vacuum degassing unit, shortly
• Filter samples before use in a 1200 Infinity Series system. Use High pressure
• When using buffer solutions, flush the system with water before switching it
• Check the module plungers for scratches, grooves and dents when changing
• Solvent Information - Observe recommendations on the use of solvents, see
• Priming and Purging the System - When the solvents have been exchanged or
Table 11
Activity
After an installation
When switching between
reverse phase and normal
phase (both times)
After an installation
To clean the system when
using buffers
After a solvent change
Agilent InfinityLab LC Series Vialsamplers User Manual
degas your samples before using them in the module.
filter kit (5067-4638) for inline filtering.
off.
the piston seal. Damaged plungers cause micro leaks and will decrease the
lifetime of the seal.
"Solvent Information"
the system has been turned off for a certain time (for example, overnight)
oxygen will re-diffuse into the solvent channel. Therefore priming and purging
of the system is required before starting an application.
Choice of Priming Solvents for Different Purposes
on page 108.
Solvent
Isopropanol
Isopropanol
Ethanol or methanol
Bidistilled water
Bidistilled water

Preparing the Module

Preparing the Module
Comments
Best solvent to flush air out of
the system
Best solvent to flush air out of
the system
Alternative to isopropanol
(second choice) if no
isopropanol is available
Best solvent to re-dissolve
buffer crystals
Best solvent to re-dissolve
buffer crystals
5
107