7
Troubleshooting and Diagnostics
Maintenance Functions
Next
Remove the gripper adapter.
Next
Remove calibration fixture.
Back
Completes the procedure.
Vialsampler Gripper Verification Test
Gripper Verification Description
The verification procedure uses several vial positions as reference points to
verify the gripper alignment is correct. If verification indicates one or more
positions are out of alignment, the Automatic Referencing procedure should
be done.
Allows you to select a vial position. You can select from positions 1, 11, 39, 56,
66, 101, 111, 139, 156 and 166.