Overview Of The Module's Indicators And Test Functions - Agilent Technologies InfinityLab LC Series User Manual

Vialsamplers
Hide thumbs Also See for InfinityLab LC Series:
Table of Contents

Advertisement

7
Troubleshooting and Diagnostics

Overview of the Module's Indicators and Test Functions

Overview of the Module's Indicators and Test Functions

Status Indicators
The module is provided with two status indicators which indicate the
operational state of the module. The status indicators provide a quick visual
check of the operation of the module.
Error Messages
In the event of an electronic, mechanical or hydraulic failure, the module
generates an error message in the user interface. For each message, a short
description of the failure, a list of probable causes of the problem, and a list of
suggested actions to fix the problem are provided (see chapter Error
Information).
Maintenance Functions
The maintenance functions position the needle arm, gripper assembly, and
metering device for easy access when doing maintenance (see
Functions"
Tray Alignment
Tray alignment is required after repair of internal components. The procedure
aligns the gripper arm correctly to ensure the positioning of the gripper arm is
correct for all vials (see
142
on page 144).
"Automatic Referencing"
Agilent InfinityLab LC Series Vialsamplers User Manual
"Maintenance
on page 151).