Maintenance and Repair

Exchange the Gripper Arm

Exchange the Gripper Arm

When
Defective gripper arm
Tools required
Description
Straightened paper clip.
Parts required
p/n
G1313-60010
Preparations
Finish any pending acquisition job
1
In the Local Controller start the maintenance mode and
select Change Gripper function.
OR
In the Tools section of the Agilent Lab Advisor software
select Service & Diagnostics > Maintenance Positions
> Change Gripper, click Start and wait until the gripper
is in maintenance position.
2
Open the front door.
216
Description
Gripper assembly
3
Identify the slit below the gripper motor and the gripper
arm release button.
Agilent InfinityLab LC Series Vialsamplers User Manual