Change Metering Seal or Piston

The change-metering seal/piston function draws the piston away from the
home position, relieving the tension on the spring. In this position, the
analytical head assembly can be removed and reinstalled easily after
maintenance.
User Interface
The commands for the control software are:
Start
Draws the piston away from the home position, relieving the tension on the
spring.
Next
Repositions the piston at the home position.
Read carefully the description in the Lab Advisor user interface.
N O T E
Back
Completes the procedure.
Using the Change Seal Function
1 Select Start to move the piston to the maintenance position.
2 Exchange the metering seal (see
3 Select Next to move the piston back to the home position.
Read carefully the description in the LabAdvisor user interface.
N O T E
4 Select Back to finish the procedure.
Agilent InfinityLab LC Series Vialsamplers User Manual
Piston"
on page 212).
Troubleshooting and Diagnostics
Maintenance Functions
"Exchange the Metering Seal and
7
147