Agilent Technologies InfinityLab LC Series User Manual page 52

Vialsamplers
Hide thumbs Also See for InfinityLab LC Series:
Table of Contents

Advertisement

2
Site Requirements and Specifications
Specifications of the Integrated Column Compartment
Performance Specifications Agilent InfinityLab LC Series
Integrated Column Compartment (G7130A)
Table 10
Type
Temperature range
Column capacity
Temperature stability
Temperature accuracy
Warm up time
52
Performance Specifications Agilent InfinityLab LC Series Integrated Column
Compartment (G7130A)
Specification
5 °C above ambient to 80 °C
2 columns up to 30 cm and
4.6 mm ID
±0.10 °C
±0.8 K (±0.5 K with calibration)
20 – 40 °C in 5 min
Agilent InfinityLab LC Series Vialsamplers User Manual
Comments
at sensor
measured with water at
ambient temperature (25 °C),
set point at 40 °C and a stable
flow of 1 mL/min at outlet of
pre-heater
at sensor

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents