Error Information
General Error Messages

Leak Sensor

Error ID: 0064
A leak was detected in the module.
The signals from the two temperature sensors (leak sensor and
board-mounted temperature-compensation sensor) are used by the leak
algorithm to determine whether a leak is present. When a leak occurs, the leak
sensor is cooled by the solvent. This changes the resistance of the leak sensor
which is sensed by the leak-sensor circuit on the main board.
Probable cause
1
2
3
4
5
The leak sensor in the sampler is hidden under the ICC Column Heater or Column Shelf
N O T E
respectively.
168
Loose fittings.
Broken capillary.
Leaking rotor seal or needle seat.
Defective metering seal.
Leaking peristaltic pump
Suggested actions
Ensure all fittings are tight.
Exchange defective capillaries.
Exchange the rotor seal or seat capillary.
• Exchange the metering seal.
• Make sure the leak sensor is thoroughly dry
before restarting the autosampler.
Exchange the peristaltic pump.