Fan Failed

Error ID: 0068
The fan in the autosampler module or in the sample cooler or sample
thermostat has failed.
The hall sensor on the fan shaft is used by the main board to monitor the fan
speed. If the fan speed falls below a certain limit for a certain length of time,
the error message is generated.
This limit is given by 2 revolutions/second for longer than 5 seconds.
Depending on the module, assemblies (e.g. the lamp in the detector) are
turned off to assure that the module does not overheat inside.
Probable cause
1
2
3
4
Agilent InfinityLab LC Series Vialsamplers User Manual
Fan cable disconnected.
Defective fan.
Defective sample cooler or sample
thermostat fan
Defective main board.
Error Information
General Error Messages
Suggested actions
Please contact your Agilent service
representative.
Please contact your Agilent service
representative.
Replace the sample cooler or sample
thermostat.
Please contact your Agilent service
representative.