Leak and Waste Handling

Leak and Waste Handling

The Agilent InfinityLab LC Series has been designed for safe leak and waste
handling. It is important that all security concepts are understood and
instructions are carefully followed.
The solvent cabinet is designed to store a maximum volume of 8 L solvent. The
maximum volume for an individual bottle stored in the solvent cabinet should
not exceed 2 L. For details, see the usage guideline for the Agilent Infinity II
Solvent Cabinets (a printed copy of the guideline has been shipped with the
solvent cabinet, electronic copies are available on the Internet).
All leak plane outlets are situated in a consistent position so that all Infinity
and Infinity II modules can be stacked on top of each other. Waste tubes are
guided through a channel on the right hand side of the instrument, keeping the
front access clear from tubes.
The leak plane provides leak management by catching all internal liquid leaks,
guiding them to the leak sensor for leak detection, and passing them on to the
next module below, if the leak sensor fails. The leak sensor in the leak plane
stops the running system as soon as the leak detection level is reached.
Solvent and condensate is guided through the waste channel into the waste
container:
• from the detector's flow cell outlet
• from the Multisampler needle wash port
• from the Sample Cooler or Sample Thermostat (condensate)
• from the pump's Seal Wash Sensor (if applicable)
• from the pump's Purge Valve or Multipurpose Valve
28
Agilent InfinityLab LC Series Vialsamplers User Manual