Integrated Column Compartment (ICC) Heater Error Messages
ICC Temp Wait Timeout Error
Error ID: 30820
The Integrated Column Compartment (ICC) Heater has detected that the left
heater control has reached a time limit.
Probable cause
1
2
ICC Heater Overcurrent Failure
Error ID: 30821
The Integrated Column Compartment (ICC) Heater has detected an
overcurrent in the heater circuits. For safety reason heater over-current
protection became active.
Probable cause
1
2
Agilent InfinityLab LC Series Vialsamplers User Manual
Integrated Column Compartment (ICC) Heater Error Messages
Defective cable or connector
Defective ICC heater
Defective cable or connector
Defective ICC heater
Error Information
Suggested actions
Please contact your Agilent service
representative.
Replace the ICC heater.
Suggested actions
Please contact your Agilent service
representative.
Replace the ICC heater.
8
183