Error Information
Pump Error Messages
Pump drive stroke blocked
Error ID: 22441
The pump drive movement is blocked.
• Parameter: 1 – 4 referring to pump drive
Probable cause
1
Pump drive stop not found
Error ID: 22442
The pump drive stop has not been found.
• Parameter: 1 – 4 referring to pump drive
Probable cause
1
Pressure sensor calibration wrong or missing
Error ID: 22443
Pressure sensor calibration wrong or missing.
• Parameter: None
Probable cause
1
98
Pump drive motor is mechanically blocked.
Pump drive spindle is defect.
Pressure sensor calibration wrong or
missing.
Suggested actions
Please contact your Agilent service
representative.
Suggested actions
Please contact your Agilent service
representative.
Suggested actions
• Replace pressure sensor.
• Please contact your Agilent service
representative.
Agilent 1290 Infinity Binary Pump User Manual