Removing The Pump Head Plunger - Waters ACQUITY UPLC Overview And Maintenance Manual

Quaternary solvent manager
Table of Contents

Advertisement

Figure 3–90: Support-plate bolts on accumulator pump head
1
Support-plate bolts
Figure 3–91: Removing accumulator pump head and support plate from actuator
Support plate
Pump head
Plunger
8.
Stand the pump head upright on a clean surface.

3.19.3 Removing the pump head plunger

Warning:
working with hazardous materials. Consult the Material Safety Data Sheets regarding
the solvents you use. Additionally, consult the safety representative for your
organization regarding its protocols for handling such materials.
Notice:
powder-free gloves when performing this procedure.
housing
Observe Good Laboratory Practice (GLP) at all times, particularly when
To avoid contaminating system components, wear clean, chemical-resistant,
August 8, 2016, 715005050 Rev. C