Waters ACQUITY UPLC Overview And Maintenance Manual page 103

Quaternary solvent manager
Table of Contents

Advertisement

Figure 3–80: Inlet manifold outlet-tubing fitting
3
Inlet manifold outlet-tubing fitting
Inlet manifold
2
i
Valve
6.
Reattach the outlet-tubing fitting to the inlet manifold, and tighten it with your fingers to the
extent possible.
Requirement:
Ensure that the outlet-tubing is fully engaged by raising the inlet manifold
while attaching the outlet-tubing fitting.
Figure 3–81: Reattaching outlet-tubing fitting
Outlet-tubing fitting
Inlet manifold
7.
Tighten the outlet-tubing fitting, on the side of the i
the 1/4-inch open-end wrench to tighten the fitting an additional 1/6-turn, for existing
fittings, or 3/4-turn, for new fittings.
1
2
1
2
August 8, 2016, 715005050 Rev. C
Page 103
2
Valve with your fingers, and then use