Figure 3–59: Lifting pump head from support plate
Pump head
Seal-wash housing
Support plate
4.
Remove the old seal-wash seal from the seal-wash housing, and discard the old seal.
Figure 3–60: Location of seal-wash seal
Seal-wash seal
Seal-wash housing
5.
Using the smooth end of the seal removal tool, remove the plunger-seal spacer from the
pump head.