Waters ACQUITY UPLC Overview And Maintenance Manual page 64

Quaternary solvent manager
Table of Contents

Advertisement

Requirement:
this procedure.
Required tools and materials
• Chemical-resistant, powder-free gloves
• 1/4-inch open-end wrench
• 5/16-inch open-end wrench
• 1/2-inch open-end wrench
• Accumulator check-valve assembly
To replace the accumulator check valve:
1.
Flush the solvent manager with nonhazardous solvent.
2.
Power-off the solvent manager.
3.
Using the 5/16-inch open-end wrench to hold the check valve in place, disconnect the
compression fitting by using the 1/4-inch open-end wrench.
Figure 3–25: Compression fitting on check valve
Place the 5/16-inch open-end wrench here
Compression fitting
Notice:
which is normally on the top face of the check valve, does not remain in the head.
4.
Use the 1/2-inch open-end wrench to loosen the check valve, and then remove the check-
valve assembly from the pump head.
Wear clean, chemical-resistant, powder-free gloves when performing
1
2
When you remove the valve assembly, ensure that the PEEK washer,
August 8, 2016, 715005050 Rev. C
Page 64

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents