EN - ENGLISH
INSTRUCTIONS FOR USE
Generator Setup
Place Generator
1
2
Plug In Generator
REAR PANEL
Foot Pedal & Electrosurgical Irrigator Setup
1
Plug In Foot Pedal
REAR PANEL
Codman Electrosurgical Irrigator Setup (Optional)
2
IRRIGATOR: REAR PANEL
Codman
Electrosurgical Generator
®
CUT
COAG
W
W
MONOCUT
BICUT
MONOCOAG
BICOAG
Monopolar
Bipolar
FOOT PEDAL
GENERATOR: REAR PANEL
Place the Generator on a stand, shelf, or other flat surface
capable of securely supporting the device.
a
Plug the power cord into the rear panel of the Generator.
b
Plug the Generator into a grounded electrical outlet of
the appropriate voltage rating.
If you are using the Foot Pedal with the Generator:
Insert the plug of the Foot Pedal into the Foot Pedal
a
Receptacle on the rear of the Generator. Place the Foot
Pedal in an area accessible to the user.
Twist the cable collar 1/4 turn to lock the Foot Pedal cable
b
in place.
Note: Verify activation of the Foot Pedal before each use.
If you are using the Electrosurgical Irrigator with the
Generator:
a
Follow the Instructions for Use provided with the Irrigator.
b
Using the Electrosurgical Interconnecting Cable, connect
one end of the cable to the Generator receptacle on the
rear of the Irrigator. This cable allows you to control
irrigation while cutting or coagulating when using the
Foot Pedal to activate the system.
c
Connect other end of the cable to the receptacle on the
rear of the generator.