Led surgical headlight system

Advertisement

Integra
®
OPERATION AND SERVICE MANUAL
LED Surgical Headlight System

Advertisement

Table of Contents
loading

Summary of Contents for Integra 90520

  • Page 1 Integra ® OPERATION AND SERVICE MANUAL LED Surgical Headlight System...
  • Page 2 Integra ® LED Surgical Headlight System...
  • Page 3: Table Of Contents

    Recharging the Battery ....................................12 Maintenance and Cleaning ..................................... 13 Troubleshooting ....................................... 14 Specifications ........................................15 Electromagnetic Compatibility (EMC) ................................16 Integra Surgical Headlight Systems Ordering Information ........................18 ® Replacement Parts & Accessories Ordering Information .......................... 19 Repair and Return ......................................20...
  • Page 4: Indications For Use

    This device is intended to be used by professional users. Contraindications The Integra LED Surgical Headlight System is contraindicated for use in neonate transillumination, ophthalmic procedures, or with photosensitive patients or patients who have received photosensitizing agents (hematoporphyrin derivatives) within three months.
  • Page 5: Table Of Symbols

    Table of Symbols No naked flames Do not dispose Listed by ETL Recycle symbol Eligible for Call2Recycle Program UN approved container Part Number Recycling symbol (Taiwan) Serial Number Restriction of Hazardous Substances (China) 5 Year safe use period In this manual, this symbol and “Warning”...
  • Page 6: General Warnings

    • The LED Surgical Headlight System must be used with the • A dead battery may result if the batteries are left uncharged recommended Integra battery or with recommended Integra or unused for a long period of time. power cords with plugs inserted into properly grounded AC power outlets.
  • Page 7: Battery Warnings

    System. If the battery shows any signs of damage, leakage, liquids. The case is not sealed. or cracking, it must be replaced immediately and only with an Integra battery recommended for or supplied with • Do not open the charger or power supply case. No user the system.
  • Page 8: Overview

    Duct System battery system. Headlight Headlight Module Cable The Integra LED Surgical Headlight System utilizes an LED light source with an active cooling 90500W system. A thermostatically- controlled cooling fan draws air at a very low flow rate through...
  • Page 9 90520 / 90520W LED Headlight w/Batteries and AC/DC Power Supply The deluxe model includes components for both the portable battery operation as well as AC/DC Power operation. • LED Headlight with Cable • 2 Batteries • Battery Holster • Battery Charger Base (comes w/ dual bay) •...
  • Page 10: Preparing To Use The System

    Integra ® LED Surgical Headlight System Preparing to Use the System- With Battery Option 1. Clip the Battery Holster in a comfortable location. 2. Insert and fully seat the Battery into the Battery Holster. For proper battery orientation, align the battery so that the 5-pin connector in the holster mates with the 5 slots on the battery.
  • Page 11 Preparing to Use the System- With Battery Option 1. Place the LED Headlight on the head, with the light module located between the eyes (Figure 5). Optimal coaxial (shadow-free) positioning is achieved when a line drawn between the pupils bisects the headlight module lens.
  • Page 12: Battery Charge Status And Low Battery Notification

    Integra ® LED Surgical Headlight System Battery Charge Status and Low Battery Notification 1. The Battery utilizes a 5-bar LCD “fuel gauge” to indicate its charge status (Figure 9): 5 bars displayed 81% - 100% charged 4 bars 61% - 80%...
  • Page 13: Recharging The Battery

    Recharging the Battery 1. Place the Battery Charger on a flat, level surface away from sources of heat and moisture. 2. Connect the Battery Charger base to the 60W power supply. Insert the supplied AC Mains cord into the power supply. Plug the AC Mains cord into any standard 100 VAC to 240 VAC 47-63 Hz outlet.
  • Page 14: Maintenance And Cleaning

    Integra ® LED Surgical Headlight System Maintenance and Cleaning Maintenance The Integra LED Surgical Headlight System requires no routine maintenance. For servicing and/or repair, refer to the ® Repair and Return section of this manual. Cleaning 1. Unplug the unit prior to cleaning.
  • Page 15: Troubleshooting

    2 hours Cooling fan is making a whirring or Fan may have been damaged after Return to Integra for service clicking sound dropping headlight The operating temperature of the LED is continuously monitored and the fan speed is adjusted to maintain proper LED operating temperatures.
  • Page 16: Specifications

    Integra ® LED Surgical Headlight System Electromagnetic Compatibility Portable LED Headlight Type Wattage 13 Watts Lamp Life >25,000 hours Color Temperature Average 4500 (minimum of 4300) Spot Size Range 25mm to 125mm at 16 in (~40cm) working distance Power Input...
  • Page 17: Electromagnetic Compatibility (Emc)

    Guidance and Manufacturer’s Declaration – Immunity All Equipment and Systems The Integra LED Surgical Headlight System is intended for use in the electromagnetic environment specified below. The customer or user should insure that it is used in such an environment.
  • Page 18 Table 204 Guidance and Manufacturer’s Declaration – Electromagnetic Immunity The Integra LED Surgical Headlight System is intended for use in the electromagnetic environment specified below. The customer or user should insure that it is used in such an environment. Immunity Test...
  • Page 19: Integra Surgical Headlight Systems Ordering Information

    Surgical Headlight Systems Ordering Information ® REF # Description Headlight Battery Kit AC/DC Power Kit 90520 Integra LED Surgical Headlight System with Batteries and AC/DC Power 90520US LED Headlight w/Batteries & AC/DC Power Supply - US 90500 90530US 90550US 90520EU LED Headlight w/Batteries &...
  • Page 20: Replacement Parts & Accessories Ordering Information

    Integra ® LED Surgical Headlight System Replacement Parts & Accessories Ordering Information 90524US Dual Bay Charger Kit - US Contains 90524 Charger, 407481 60W Charger Power Supply, 90590US Power Cord 90530US Battery Kit - US Contains 90521 Holster, 90524 Dual Bay Charger, 407481 60W Charger Power Supply, 90522 Batteries (QTY.2), 90590US Power Cord...
  • Page 21: Repair And Return

    (3) years from the date of shipment, except the batteries, which are warranted for one (1) year from the date of shipment. Integra’s sole and exclusive liability under the warranty shall be, at Integra’s option, either to repair or replace any component which fails during the warranty period due to any defect in...
  • Page 22 U.S. Patent 8,899,744; additional patent(s) pending Integra and the Integra logo are registered trademarks of Integra LifeSciences Corporation or its subsidiaries in the United States and/or other countries. ©2020 Integra LifeSciences Corporation. All rights reserved. Printed in the U.S.A.