Integra Codman 901001ESUO Instruction Manual page 15

Electrosurgical generator
Table of Contents

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 1
TECHNICAL SPECIFICATIONS AND ENVIRONMENTAL REQUIREMENTS
Current Mode
Current Type
MONOCUT (Cutting) max. 80 W ±20 % on 600 Ω
MONOPOLAR
MONOCOAG (Coag.) max. 80 W ±20 % on 400 Ω
BICUT (cutting)
BIPOLAR
BICOAG (coag.)
Classification
Mains Supply
Power Consumption
Protection Class
LF and RF Leakage Currents
Application Type
Operating Mode
Mains Fuse
Signal Level
Weight
Dimensions
Standards
Environmental
Temperature
Relative Humidity
(non-condensing)
Atmospheric Pressure
RF Output Power
max. 80 W ±20 % on 300 Ω
max. 60 W ±20 % on 50 Ω
100-240 VAC ; 50/60 Hz
Without HF output: approximately 50 VA
At maximum output power: approximately 500 VA
I
According to IEC 60601-2-2
CF; defibrillator safe
Intermittent 10 seconds ON /30 seconds OFF
2 x T 3.15 AH 250 V 5x20 mm
RF indicator: 40-65 dB(A) adjustable. Alarm: 65 dB(A)
Approximately 5.0 kg
Width 403 mm
Height 170 mm
Depth 313 mm
IEC 60601-1:2005 + A1:2012, Ed. 3.1
CAN/CSA-C22.2 No. 60601-1:14
ANSI/AAMI ES60601-1:2005/(R)2012 and A1:2012, C1:2009/(R)2012 and A2:2010/(R)2012
IEC 60601-1-2:2007, Ed. 3.0
IEC 60601-1-2:2014, Ed. 4.0
IEC 60601-2-2:2017, Ed. 6.0
Operational
10° to 40°C / 50° to 104°F
30% to 75%
700 hPa to 1050 hPa
RF Output
RF Output
Current
Voltage
0.51 A on 600 Ω max 800 Vp
0.51 A on 400 Ω max 560 Vp
0.6 A on 300 Ω
max 400 Vp
1.2 A on 50 Ω
max 440 Vp
Transport
-40° to 60°C / -40° to 140°F
15% to 85%
500 hPa to 1050 hPa
15
EN - ENGLISH
Nominal
Modulation
4.00 MHz
33 kHz
4.00 MHz
33 kHz
4.00 MHz
33 kHz
4.00 MHz
33 kHz
Storage
-40° to 60°C / -40° to 140°F
15% to 85%
500 hPa to 1050 hPa

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents