1. Find the edge.
a. Select the Search tab.
b. Select the down arrow after the Label field, and select the label
c. Select the Value field and enter 1.
d. Select the Next button to locate the edge transition.
2. Place a marker on the edge.
Select the Set G1 button. This sets global marker G1 at the location of the
edge you just found.
3. Search for the pattern. Searches start at your current location. Since you
just set the global marker G1, it indicates where the search starts from.
a. Select the down arrow after the Label field, and select the label
b. Select the Value field and enter the pattern you are searching for.
c. Select the down arrow after the When field and select Entering.
d. Select the Next button to find the next occurrence of that pattern after
If the logic analysis system cannot find the pattern, a "Value not found"
message pops up.
4. Place a marker on the pattern.
Select the Set G2 button. This will set global marker G2 at the location of
the pattern.
5. Find the time between the edge and the pattern.
a. Select the Markers tab.
b. In the G2 row, select the down arrow after from, and select G1.
The value after the from field changes to the time between G1 and G2. You
can toggle between time and samples by selecting the arrow after the Time
or Samples field.
See Also
Using the Digital Waveform Display Tool (see the Waveform Display Tool
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Chapter 1: Agilent Technologies 16712A 128K Sample Logic Analyzer
containing the edge.
containing the pattern.
G1.
Interpreting the Data
27