Agilent Technologies 16712A Help Manual page 101

128k sample logic analyzer
Table of Contents

Advertisement

Chapter 1: Agilent Technologies 16712A 128K Sample Logic Analyzer
The Trigger Tab
Selecting the Arming Control... button brings up the Machine
Arming Tree dialog. In this dialog you can set what starts each
analyzer, and which one sends a signal to Arm Out. Arm Out is used to
send signals to other instruments in the frame, or sent to Port Out.
Port Out can be used to control additional instruments external to the
logic analysis system.
To change the source of Arm In or the destination of Port Out, use the
Intermodule dialog (see the Agilent Technologies 16700A/B-Series
Logic Analysis System help volume).
Setting One Analyzer to Arm the Other
This procedure assumes you have both analyzers turned on. The
second analyzer can be turned on by dragging it onto the workspace in
the Workspace window.
1. Under the Trigger tab, select the Settings subtab.
2. Select the Arming Control... button.
3. Select the box of the machine that you want to have wait for its arm signal.
4. In the armed by menu, select the other machine.
The arming signal path forms a tree in the Machine Arming Tree dialog.
5. Set the sequence level number to the trigger sequence level that will wait
for the arm signal. It does not have to be the same as the trigger level.
6. Select the Close button.
NOTE:
If the trigger sequence does not pass through the level containing the wait for
arm term, the trigger will not wait for the arming signal.
See Also
Overview - Multiple Instrument Configuration (see the Agilent
Technologies 16700A/B-Series Logic Analysis System help volume)
Overview - Multiple Machine Configuration (see the Agilent Technologies
16700A/B-Series Logic Analysis System help volume)