The Sampling Tab

The options under Sampling tell the analyzer the overall way in which
you want to make a measurement. The options include:
• The acquisition mode.
• The data sample rate.
• How much data before and after the trigger.
• How much data to acquire in all.
See Also
"Naming the Analyzer" on page 45
"Turning the Analyzer Off" on page 46
"Setting the Acquisition Mode" on page 42
"Sample Period (Timing Only)" on page 46
"Performing Clock Setup (State only)" on page 42
"Trigger Position Control" on page 51
"Acquisition Depth" on page 41

Acquisition Depth

Acquisition Depth, located under Sampling and also under Trigger
Settings, sets how much data will actually be acquired. While the
Agilent Technologies 16712A logic analyzer has a maximum memory
depth of 128 K samples in State Mode and 256K in Timing Mode,
sometimes you may not want to wait for all that memory to fill up.
The numbers shown in the Acquisition Depth menu are
approximations. The combination of count tags, pod assignments, and
acquisition modes affect what choices are available. Also, the values
are memory depth per channel.