To Assign Bits
1. Select the bit assignment label to the right of the label name you want to
define.
Each bit assignment label corresponds to the data pod which is listed
above it.
2. Either choose one of the predefined groups, or Individual.
3. In Individual, select the bits to toggle them between an asterisk and a
period.
NOTE:
Labels can have a maximum of 32 channels assigned to them.
Bits assigned to a label are numbered from right to left. The least
significant assigned bit on the far right is numbered 0. The next
assigned bit to the left is numbered 1, and so on. Labels can contain
bits that are not consecutive; however, bits are always numbered
consecutively within a label. Above each column of bit assignment
fields is a bit reference line that shows you the bit numbers and activity
indicators.
See Also
"To reorder bits in a label" on page 67
"Activity Indicators" on page 62

Inserting and Deleting Labels

To Insert Additional Labels
1. Select the label name button that you want to insert another label next to.
2. Choose Insert before... or Insert after....
To Delete Labels
1. Select the button of the label name that you want to delete.
2. Choose Delete.
The bit assignments of deleted labels are not saved. You can make a
label inactive but not lose its assignment by turning it off (see page 66)