Chapter 1: Agilent Technologies 16712A 128K Sample Logic Analyzer
The Sampling Tab
help to move fast lines from pod 1 to pod 2; they must be moved to pod
3, which is a different pod pair.
Invalid Data
The analyzer only looks for transitions on bits that are assigned. If the
bit is assigned after a run by changing the label and selecting Apply in
the analyzer window, another run is automatically started.
Comparison of Transitional and Conventional Timing
Modes
In Conventional Timing Acquisition mode, the analyzer stores
measurement data at each sampling interval. In Transitional Timing
Acquisition mode, the timing analyzer samples data at regular intervals,
but only stores data when there is a threshold level transition. Each
time a level transition occurs on any of the bits of a pod pair, data on all
channels is stored. A time tag is stored with each stored data sample so
the measurement can be reconstructed and displayed later.
When using transitional timing, you might consider how many
transitions can be stored. The number depends on the mode and
frequency of transition occurrence.
See Also
"Timing Acquisition Modes" on page 47
"Transitional Timing Acquisition Modes" on page 47
50