Importing Netlist and ASCII Files
instead.

Turning Labels On and Off

You may want to turn off labels that you have created so that the label
is not displayed. When a label is turned off, its name and bit
assignments are preserved.
To Turn Off a Label
1. Select the button of the label name that you want to turn off.
2. Choose Label [ON] to toggle it off.
If the label is the only one, it cannot be turned off or deleted. If there is
more than one label but it is the only one on, turning it off turns the first
label back on.
To Turn On a Label
1. Select the button of the label name that you want to turn on.
2. Choose Label [OFF] to toggle it on.
To Display a Label that was Off
1. Turn on the label.
2. At the bottom of the window, select the Apply button.
The label's data appears in the display windows.

Label Polarity

The analyzer uses the label polarity to identify patterns when
triggering and displaying data.
To change the label polarity, select the polarity field, which toggles
between positive (+) and negative (-). Positive polarity means that a
high voltage is a logical "1". Negative polarity means that a high voltage
is a logical "0".
Changing the label polarity after you have set up other parts of the
measurement changes these things:
• "1" and "0" values flip in trigger terms