Timer control is not available in the first sequence level.
3. Select the Off option button at the bottom of the edit dialog box and
choose Start.
If the other analyzer is already using a timer, it won't appear in this
analyzer's edit dialog.
4. Select the event field where you want to use the timer.
5. Choose either Timer or Combo....
nanoseconds.
6. In the other trigger levels, you can Start, Pause, and Continue the timer.
When a timer reaches its assigned value, it automatically stops.
See Also
"Using Combinations of Terms" on page 97
Using Combinations of Terms
Resource terms can be used in combinations. Combinations use the
logical AND, NAND, OR, NOR, and XOR functions to combine
predefined resource terms.
A combined term is evaluated as a single events. Components that are
ANDed together must all occur in the same sample. Ones that are
ORed together only require one of the components to happen.
If you intend for the logic analyzer to decide between two actions, use
branches instead of combined terms. If you want the logic analyzer to
find the terms in sequence, use different levels or a trigger function in
the trigger sequence.
To Set Up a Combination
1. Edit the sequence step (see page 78) that you want a combination term to
appear in.
2. Select the resource term assignment field that you want to make into a
combination.
3. Choose Combo... from the menu of resource terms.
4. In the Combination dialog, select the On/Off/Not option button for each
term that you want to use and select On or Not.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Chapter 1: Agilent Technologies 16712A 128K Sample Logic Analyzer
Use Combo for cases such as an edge occurring after so many
The Trigger Tab
97