Cleaning your cycler
The Liberty Select Cycler is not sterile equipment,
however, you should keep your Liberty Select Cycler clean
and dry at all times. Clean the exterior surface of your
machine using a damp cloth and a diluted (1:100) bleach
solution or other suitable hospital disinfectant. See adjacent
illustration for areas to be cleaned.
Use surface cleaner in small amounts to avoid excess
cleaner from entering the interior of your machine. Rinse
off cleaning solutions with a water-dampened cloth,
especially if a corrosive cleaner such as bleach is used.
If you need to clean your stay•safe organizer, you may
wash it with warm soapy water or place it in the top rack of
your dishwasher.
Periodically check your Liberty Select Cycler for:
Damage to the cabinet
Loose or missing pieces
There is no other preventative maintenance for your cycler.
If your cycler needs service, call an FMC Technical
Support Representative at 1-800-227-2572.
Liberty Select Cycler User's Guide
Caution: Do not allow any fluid to drip into your
cycler or onto the pump module. The cycler must
be OFF when you are cleaning it to prevent fluid
from entering the pump module. If any fluid gets
inside your cycler, you must contact a Fresenius
Medical Care Technical Support Representative
immediately at 1-800-227-2572. Clean fluid spills
immediately to prevent moisture from possibly
damaging your machine. Your machine should be
placed on a water-tolerant surface, and spills
should be cleaned up immediately.
Caution: Do not touch the white pressure sensors
in the pump module. Pressing on the pressure
sensors may damage them.
Caution: Do not use foaming type cleansers or
disinfectants containing alcohol or quaternary
ammonium compounds like N-alkyl (C12-C18)
dimethyl benzyl ammonium chloride
(benzalkonium chloride). These chemicals attack
the polycarbonate plastics used in your machine.
Read the product labels and follow the instructions.
Do not use spray bottles.