Fresenius Medical Care Liberty Select User Manual page 187

Peritoneal dialysis cycler
Table of Contents

Advertisement

1 Press STOP
If you are performing a Fill, Dwell, or Pause, you may
choose to do a STAT (immediate) drain. This unplanned
Drain will interrupt your Fill, Dwell, or Pause. Press the
STOP key to first stop your Fill, Dwell, or Pause.
2 Touch STAT drain
After the "Stopped" screen displays the message "Press
OK key to continue fill," touch the STAT drain button.
3 Touch STAT drain on the next screen
Your cycler will display the "Select STAT drain option"
screen. To continue with the STAT drain, touch the STAT
drain button.
If you were previously in a Fill or Dwell cycle, the STAT
drain will immediately begin draining your abdomen.
If you were previously in a Pause cycle, your cycler will
ask you to reconnect first. See page 146 for information on
how to reconnect after a Pause. After you reconnect, the
STAT drain will immediately begin draining your
abdomen.
During a STAT drain, your cycler will display a "Drain"
screen with the word 'STAT' in the upper left corner to
show that it is running an unplanned drain.
After your STAT drain is complete, your cycler will either
begin a short Dwell or Pause (depending on which cycle
you interrupted) so your cycler can prepare the heater bag
for your next cycle. Afterward, your cycler will begin your
next programmed Fill, if required. Your treatment data will
show that you performed a STAT drain (see page 176 for
more information).
If you choose not to perform a STAT drain, either touch the
Cancel button or wait 10 seconds. Your cycler will go
back to the "Stopped" screen (see page 182).
Liberty Select Cycler User's Guide
Note: A STAT Drain works the same as a regular
Drain. You may stop it or bypass it as needed.
Note: An additional drain is a planned drain. STAT
drain is an unplanned drain.