9 Select values
Touch the yellow button to select the value it displays:
Tidal drain volume—How much you are to be drained
each exchange
Last fill volume—The amount of your Last Fill
(this is optional and may be set to zero per the physician
prescription)
10 Enter values
When you touch a yellow button, it will turn orange. Use
the Up/Down Arrow (/) keys to set it to your
prescribed value.
Repeat these steps to enter all of your prescribed values.
11 Press OK
Press the OK key to confirm and save your settings.
12 Touch Back
Touch the Back button to return to the previous screen,
then touch the Back button again to reach the "My
Settings" main screen.
Skip to page 76 to enter your personal Patient Settings.
Liberty Select Cycler User's Guide
Warning: Your Tidal Drain Volume must include
your Tidal Fill Volume plus your expected
ultrafiltration (UF) per cycle, which is based on an
average daily UF volume from the Tidal Dwell in
order to prevent an overfill situation. Overfilling
can cause serious injury or death. For more
information on overfilling, see page 190.
Note: Light blue buttons cannot be selected. Your
machine calculates the values in blue buttons based
on the values you enter in the yellow buttons.
Note: If you need to set the last bag option, see
"Options Settings," on page 86.
P/N 480145 Rev A
71